7. CALIBRATION

7-1 Calibration Group

The GX series balance has the following modes as a calibration group.

Calibration

Automatic self calibration (calibration due to changes in temperature)

Calibration using the internal mass (one-touch calibration)

Calibration using an external weight

Calibration test

Calibration test using an external weight (Calibration test does not perform calibration.)

Correction of the internal mass value

Terms

The following terms are defined as follows:
Internal mass

= Built-in calibration weight

External weight = A weight that you have. Referred to as a calibration weight when used for calibration.
Calibration weight = A weight used for calibration
Target weight

= An external weight used for calibration test

Caution

Calibration adjusts the balance for accurate weighing.

Besides periodic calibration and before each use, perform calibration when:
• the balance is installed for the first time.
• the balance has been moved.
• the ambient environment has changed.

Do not allow vibration or drafts to affect the balance during calibration.
To output the data for GLP using the RS-232C interface, set “GLP output ( info )” of “Data output

( dout )”. For details, refer to “9. FUNCTION TABLE”. Time and date are added to GLP report. If
the time or date is not correct, adjust them. For details, refer to “9-9 Clock and Calendar Function”.

Calibration test is available only when “GLP output ( info )” of “Data output ( dout )” is set to “1” or “2”,
The calibration and calibration test data can be stored in memory. To store them, set “Data

memory ( data )” to “3”. For details, refer to “11. DATA MEMORY”.

Caution on using an external weight

The accuracy of an external weight can influence the accuracy of weighing. Select an

appropriate weight as listed below:

Model

Usable calibration weight

Adjustable range

GX-200

200 g, 100 g

GX-300 300

g,

200 g, 100 g

GX-400

400 g, 300 g, 200 g

GX-600 600

g,

500 g, 400 g, 300 g, 200g

GX-800

800 g, 700 g, 600 g, 500 g, 400 g, 300 g, 200g

GX-1000 1000g, 900g, 800 g, 700 g, 600 g, 500 g, 400 g, 300 g,

200g

-0.015 g to +0.015 g

GX-2000 2000 g, 1000 g

GX-3000 3000

g, 2000 g, 1000 g

GX-4000 4000 g, 3000 g, 2000 g

GX-6100 6000

g,

5000 g, 4000 g, 3000 g, 2000g

-0.15 g to +0.15 g

GX-6000 6000

g,

5000 g, 4000 g, 3000 g, 2000g

GX-8000 8000 g, 7000 g, 6000 g, 5000 g, 4000 g, 3000g, 2000g

-1.5 g to +1.5 g

The calibration weight in bold type: factory setting