73

Display Error code

Description

Clock

battery

error

The clock backup battery has been depleted.

Press any key and set the time and date. The clock and
calendar function works normally as long as the AC
adapter is connected to the balance. If this error appears
frequently, contact the local A&D dealer.

(Blinking)

Memory full

The amount of weighing data in memory has reached
the maximum capacity.

Delete data in memory to store new data. For details,
refer to “11. DATA MEMORY”.


(Illuminated)

Memory full

The amount of calibration or calibration test data in
memory has reached the maximum capacity (50 sets).

The data in memory will be deleted automatically to
store new data. For details, refer to “11. DATA
MEMORY”.

Memory type error

Type of memory set in the function table and type of
data stored are different.

For details, refer to “11. DATA MEMORY”.

EC, E00

Communications error

A protocol error occurred in communications.

Confirm the format, baud rate and parity.

EC, E01

Undefined command error

An undefined command was received.

Confirm the command.

EC, E02

Not ready

A received command can not be processed.

e.g. The balance received a Q command, but not in the

weighing mode.

e.g. The balance received a Q command while

processing a RE-ZERO command.

Adjust the delay time to transmit a command.

EC, E03

Timeout error

If the timeout parameter is set to “t-Up1”, the balance
did not receive the next character of a command within
the time limit of one second.

Confirm the communication.

EC, E04

Excess characters error

The balance received excessive characters in a
command.

Confirm the command.

Advertising