2 manual response adjustment – Rice Lake GX Series, A&D Weighing User Manual

Page 21

Advertising
background image

19

Notes
If improper performance is found in the self check, the
balance displays
CH no. Contact the local A&D dealer
for repair.
If the automatic response adjustment fails, the balance
displays
CH ng. Check the ambient conditions such
as breeze and vibration, also check the weigning pan.
Then, perform the adjustment again. To return to the
weighing mode, press the CAL key.

If the automatic response adjustment is awkward, try to
refine it using the manual response adjustment.

6-2 Manual Response Adjustment

This function manually updates the response adjustment.

Operation

1 Press and hold the MODE key until RESPONSE is

displayed.

And then, press the MODE key again quickly.

2 Press the MODE key to select a rate of the response

adjustment. Either FAST, MID. or SLOW can be
selected.

3 After a few seconds of inactivity the balance displays

e

end . Then, it returns to the weighing mode and

displays the updated response indicator. The

response indicator remains displayed for a while.

Note
The response adjustment can be changed at “Condition
(
Cond)” of “Environment, Display (ba5fnc)” in the
function table. For details, refer to “9. FUNCTION
TABLE”.