2 manual response adjustment – Rice Lake GX Series, A&D Weighing User Manual
Page 21
19
Notes
If improper performance is found in the self check, the
balance displays CH no. Contact the local A&D dealer
for repair.
If the automatic response adjustment fails, the balance
displays CH ng. Check the ambient conditions such
as breeze and vibration, also check the weigning pan.
Then, perform the adjustment again. To return to the
weighing mode, press the CAL key.
If the automatic response adjustment is awkward, try to
refine it using the manual response adjustment.
6-2 Manual Response Adjustment
This function manually updates the response adjustment.
Operation
1 Press and hold the MODE key until RESPONSE is
displayed.
And then, press the MODE key again quickly.
2 Press the MODE key to select a rate of the response
adjustment. Either FAST, MID. or SLOW can be
selected.
3 After a few seconds of inactivity the balance displays
e
end . Then, it returns to the weighing mode and
displays the updated response indicator. The
response indicator remains displayed for a while.
Note
The response adjustment can be changed at “Condition
(Cond)” of “Environment, Display (ba5fnc)” in the
function table. For details, refer to “9. FUNCTION
TABLE”.