6 description of the item “data format – Rice Lake GX Series, A&D Weighing User Manual

Page 41

Advertising
background image

39

dout data

2

Data memory function is used.

Stores weighing data.

dout int Interval

time

Optional setting

dout

5-td1, 2, or 3 Adds the time and date.

Example

“For periodical weighing without a computer command and
outputting all of the data to a computer at one time”

9-6 Description of the Item “Data format”

A&D standard format

5if type 0

This format is used when the peripheral equipment can receive the A&D format. If an AD-8121B is
used, set the printer to MODE 1 or 2.

This format consists of fifteen characters excluding the terminator.

A header of two characters indicates the balance condition.

The polarity sign is placed before the data with the leading zeros. If the data is zero, the plus

sign is used.

The unit, consisting of three characters, follows the data.

DP (Dump print) format

5if type 1

This format is used when the peripheral equipment can not receive the A&D format. If an AD-8121B
is used, set the printer to MODE 3.

This format consists of sixteen characters excluding the terminator.

A header of two characters indicates the balance condition. No overload header is used.

The polarity sign is placed before the data, with spaces in place of leading zeros, if the data is

not zero or overloaded.

The unit, consisting of three characters, follows the data.

KF format

5if type 2

This is the Karl-Fischer moisture meter format and is used when the peripheral equipment can only
communicate using this format.

This format consists of fourteen characters excluding the terminator.

This format has no header characters.

The polarity sign is placed before the data, with spaces in place of leading zeros, if the data is

Advertising