Rice Lake GX Series, A&D Weighing User Manual

Page 53

Advertising
background image

51

Title block and end block

When a weight value is recorded as the GLP data, “Title block” and “End block” are inserted at the
beginning and at the end of a group of weight values, in the GLP report.

Note

To output the report to an AD-8121B , use MODE 3 of the AD-8121B.

Caution

If the data memory function is used, the “Title block” and “End block” can not be output.

Operation

1 With the weighing data displayed, press and hold the PRINT key until S5tart is displayed.

The “Title block” is output.

2 The weighing data is output according to the parameter setting of the data output mode.

3 Press and hold the PRINT key until recend . The “End block” is output.

When the setting is “info 1”:

When the setting is “info 2”:

Advertising