Rice Lake TradeRoute HL Series - Low Profile Installation and Setup for Models from 2011 and Older 128406 User Manual
Page 29
26
must not be cut closer than 6 feet away from the load cell (i.e. at least 6
feet of wire must be left attached to the load cell). The scale must be
balanced and calibrated AFTER cutting load cell wires. See Figure 19 for
where to wire the load cells to the circuit board.
5. For each load cell attach to the load cell cable according to Figure 19. The wiring
color code for the load cells is also shown.
V1
V2
V3
V4
+E
- E
+S
- S
GND
INDIC
AT
OR
LOA
DCE
LL
#1
LOA
DCE
LL
#2
LOA
DCE
LL
#3
LOA
DCE
LL
#4
RED
BLACK
GREEN
WHI TE
SHIELD
Figure 19 – Circuit Board, Load Cell Connections & Trimmer Locations
5. The indicator connection must also be completed if using a Junction box. Route
conduit and scale cable from the Junction box to the indicator.
6. Connect the scale cable to the indicator as shown below.