34

5. Repeat step 3 and 4 if necessary.

6. If temporary racks were used to load the scale, remove the racks and dead load the

scale once more.

7. The scale is now calibrated. Exit the calibration mode by entering:

16.3 Legal for Trade Inspection
The exact inspection procedures may vary slightly from those described below in your
region. The following is the basic inspection procedure:

1. Balance test (as described on page 16).

2. Increasing/Decreasing load test.

3. Out-of-Level Increasing/Decreasing load test.

4. Twist Condition Increasing/Decreasing load test.

5. Out-of-Level shut off test.

6. Printing of Ticket.