16.2 Calibration Procedure
(Refer to the M2000 manual for further details)

1. Enter the calibration mode Channel 1 on the M2000 by keying in:

The display will be showing a blinking “PASS” message. This prompts you to enter a
4 digit password. The factory default password is “1111”. This password can be
changed, but it is recommended that the factory default password be kept. You are
now in calibration mode. You will see a "C" flash on the left side of the display while

in calibration mode.

2. With the container or tanks empty, you must now dead load the scale. If

temporary racks are used to place test weights on the scale, dead load the scale with
the racks on. Ensure no product is in the container and key in:

You will see a flashing "

" on the display while the indicator is automatically dead

loading the scale. When it has completed the display will be at or near zero.

3. Now you must set the span of the system. Place a known amount of weight on the

scale. Key in the following:

Now enter the correct value of the weight followed by the TARE key. As an
example if you have 15000 lbs. of weight on the scale (the display will read an
improper value at this point) you would key in:

The display will now change to read 15000.

4. Remove the weight.