Directions for Use

Manage the AcuityLink option

55

Escalation mode

In escalation mode, a patient alarm is delivered to clinicians’ mobile devices in the order
shown below. Timing can vary, depending on selections made in the WACS Clinician
Notifier Configure Notification page.

Note

A patient’s alarms immediately escalate if the patient is unassigned for primary or
secondary notification.

1.

2.

3.

11:00:00

Primary notification

Violet Prince’s patient monitor and the Acuity Central
Station indicate that she is in an alarm state.

Seconds later, the alarm occurs on the mobile
devices of primary responders— clinicians in the
unit who have checked Violet Prince.

No one in the unit suspends her alarm tone.

11:00:30

Secondary notification

Violet’s alarm tone continues at the bedside monitor
and at the Acuity Central Station, and it escalates to
mobile devices of secondary responders— clinicians
in the unit who have checked Receive Secondary
Notification.

No one in the unit suspends her alarm tone.

11:01:00

Broadcast notification

Violet’s alarm tone continues at the patient monitor
and at the Acuity Central Station, and it broadcasts
to all active mobile devices in the unit.