Manage the AcuityLink option
59
Specify patient alarm notification to mobile devices running Clinician Notifier
1.
In the New Value column for the item that you are configuring, take the appropriate
step.
•
Pull-down menus: Click the pull-down menu and make a selection.
•
Fields: Take these steps:
a.
Hover your pointer over a field in the row.
The range of acceptable values appears.
b.
In the field, type a value within the acceptable range of values.
2.
If you have completed all configurations in the Configure Notification page, take these
steps:
a.
Click Save Changes.
Your changes are saved.
b.
Read the instructions in the row of the item(s) that you have configured, and then
take one or both of the following steps, as specified:
•
Restart the Clinician Notifier program on all mobile devices.
•
Reboot the WACS computer. For instructions, see
The changes are implemented.
WARNING If you turn on audio alarm only for Clinician Notifier event delivery,
vibratory notification does not occur at the mobile device, even if the clinician
uses the mobile device operating system to turn on vibration.