Compur statox 4120 – Compur STATOX 4120 User Manual

Page 12

Advertising
background image

Compur Statox 4120





Example of cable specifications:

Max. cable capacity:

l x C

L

= C

o

- C

i

= 107 nF

Max. cable inductivity:

l x L

L

= L

o

- L

i

= 6,6 mH

Diameter:

> 0,75 mm

2

The Statox 4120 Systems will work reliably if the electrical resistance is : R

L

< 50

Ω.

Most commonly the maximum length is limited by the cable capacity.

Example:
Cable diameter is 1,0 mm

2

(0,04 in), C

L

= 90 pF/m, L

L

= 0.7mH/km, R

L

/ km = 19,5

Ω.

The maximum cable length results to be
l =

m

pF

Ci

Co

/

90

= 1189 m (3900 ft), the electrical resistance is

R

L

= 2 x 1.189km (0,738 miles) x 19,5

Ω/km = 46,4Ω. So R

L

< 50

Ω is OK.




Sensor head connection procedure:

Open sensor head, remove cable gland.

Connect signal cable to terminals X3/1 und X3/2, polarity does not matter.

(DIN EN 60079-14 chapter 12 is applicable)

Fasten cable gland.
Connect battery pack: - Blue to terminal X4/1

- Orange to terminal X4/2

- Red to terminal X4/3

Connect signal cable to backplane of the rack. (See chapter 3.2)

pic.9: Sensor head pcb terminals



12