Compur statox 4120 – Compur STATOX 4120 User Manual

Page 29

Advertising
background image

Compur Statox 4120





9.

The option zero adjustment


This option is available for COCl

2

sensor heads with measuring range 0.3 ppm only. The zero

adjustment of the sensor is started via diagnostic box.

System requirements:
Sensor head EPROM index 18c, control module EPROM index 03a, diagnostic box EPROM index 02.

A zero adjustment

must only be done in clean air. It is started with the buttons ZERO + ENT. It

should be done before starting a calibration.

The set zero value is not visible.

If no zero adjustment has been done ever, the default value is 0.

After replacing a control module, a new zero setting has to be carried out. The zero value is
stored in the control module and is transmitted to the sensor head.

An error message C1 caused by a too high sensor zero current will only be cancelled after a
successful self test.







































29