Table 2: switch position for hcn, Compur statox 4120 – Compur STATOX 4120 User Manual

Page 16

Advertising
background image

Compur Statox 4120





DIP switch S1 (see pic. 8) programs gas and measuring range. If you change your Statox to another
gas or measuring range, a hardware modification may be required and the electronics must be re -
adjusted. This must be done by authorised personnel.


Table 1: Switch position for H

2

S

Measuring range:

0-30 ppm

0-50 ppm

0-100 ppm

S1/1 ON

OFF

OFF

S1/2 ON

ON

ON

S1/3 ON

ON

ON

S1/4 ON

ON

OFF

S1/5 ON

ON

OFF

S1/6 ON

OFF

ON

S1/7 ON

ON

OFF

S1/8 ON

ON

ON

E-PROM Index sensor head

12, 13, 14

12, 13, 14

12, 13, 14

E-PROM Index control module

≥ 02

≥ 02

≥ 02




Table 2: Switch position for HCN

Measuring range:

0-15 ppm 0-30 ppm 0-50 ppm 0-100 ppm

S1/1 ON

ON

OFF

OFF

S1/2

ON ON ON ON

S1/3

ON ON ON ON

S1/4 OFF

OFF

OFF

ON

S1/5

OFF ON ON ON

S1/6

OFF ON OFF ON

S1/7

OFF ON ON ON

S1/8

ON ON ON ON

E-PROM Index sensor head

13

12, 13, 14

12, 13, 14

15

E-PROM Index control module

≥ 02

≥ 02

≥ 02

≥ 02





Table 3: Switch position for COCl

2

Measuring range:

0-0,1 ppm

0-0,3 ppm

0-0,3 ppm 0-0,5 ppm 0-1,5 ppm

0-15 ppm

0-100 ppm

S1/1

OFF ON ON OFF OFF OFF OFF

S1/2

ON ON ON ON ON ON ON

S1/3

ON ON ON ON ON ON ON

S1/4 OFF

ON

ON

ON OFF ON ON

S1/5

ON OFF ON OFF ON OFF OFF

S1/6 ON

ON

ON

OFF OFF OFF ON

S1/7 ON

ON

ON

ON OFF OFF ON

S1/8

ON ON ON ON ON ON ON

E-PROM Index
sensor head

20

12, 13, 14

18

12, 13, 14

12, 13, 14

12, 13, 14

17

E-PROM Index
control module

04

02

02

02

02

02

02




16