Compur statox 4120, 4 diagnostic box – Compur STATOX 4120 User Manual

Page 9

Advertising
background image

Compur Statox 4120





LED A1 (9)

- On if pre - alarm threshold is exceeded

Display A2 (10)

- On if main alarm threshold is exceeded

Button A1 (4)

- Press and hold button A1 to display A1 threshold. Adjust with Pot. (12). A1 must be < A2.

Button A2 (3)

- Press and hold button A2 to display A2 threshold. Adjust with Pot. (11).

Button T (5) „Test“

- Initiates self test.

Button R (6) „Reset“
- Rests Alarms A1, A2 und SF provided the reason for the alarm has disappeared. (If SF has
been triggered by a not passed self test, only a successful self test will reset it).




2.4 Diagnostic box

pic.5: Diagnostic box

The diagnostic box must not be charged in classified areas or with other than charger Art.# 518850.
It has the following purposes:

Calibration
Fault diagnosis
On site display
Function test



9