Compur Statox 4120





Slide switch S1:

S1

ON

OFF

1 Relay active during alarm A2

Relay inactive at alarm A2

2 Relay active during alarm A1

Relay inactive at alarm A1



Toggle switch S3:

S3 ON

OFF

1

Detection mode

-

2

Rest Alarm by resetbutton R

auto - reset

3

Detection mode

-

4

Detection mode

-

5

Measured value accepted if logic
signal is 0V

-

6

Detection mode

-

7

Detection mode

Modem - and analog signal
adjustment for service only!

8

Detection mode

Delete EPROM for service only!

Ex works settings






3.4 Switch positions of the sensor head pcb


DIP switch S2 (see pic. 9) programs the following parameters:

S2

ON

OFF

1 -

Detection mode

2 Measured value in ppm

Measured value in TLV

3 No self test after “power on“

Self test after “power on“

4 Cal. Factor =1

Factor information from control
module

5 No self test

Self test in a 24 h interval

6 -

-

7 Detection mode

-

8 Detection mode

Modem “On” for service

Ex works setting