7. The maintenace request option


Purpose:

This option can differentiate between fatal errors in the sensor head and maintenance requirements. It
requires a Statox 4120 control module equipped with EPROM Index ≥ 03. It also sets the analog
output to 4 mA during the self test, if required.


Operation modes:

The Statox 4120 can operate in 9 different modes. The relevant mode can be seen by the control
module LED`s and relays.


Operation mode Statox

Analog output

SR-LED

Control
module

SF-LED

Control
module

SF-Relay MR-

Relay

PCS-*

Relay

O

Self test

0mA / 4mA / 0V

„on“

„on“

active

passive

active

O

Diagnostic box connected:

0mA / 4mA / 0V

flashing „off“ active

passive

active

O

Alarm threshold adjustment:

0mA / 4mA / 0V „on“

„off“ active passive

active

O

Start-mode

4mA / 0V

flashing

„off“

active

passive

passive

O

Detection mode:

Analog signal

„on“

„off“

active

passive

passive

O

Maintenance request:

B1 Battery Low
B2 Battery low under load
C1 Purge before test
C2 Purge between test peaks
C3 Sensitivity
(programmable by S3)
C4 Purge
PU Pump current too high

T2 Temperature out of specs






Analog signal






flashing






„off“






active






active






passive

O

System failure:

A1 Amplifier
C3 Sensitivity
(programmable by S3)


4mA / 0V


„off“


„on“


open


passive


passive

O

Data transmission error

4mA/0V

„off“

„on“

open

passive

passive

O

Power failure

0mA/0V

„off“ „off“

open

passive

passive


*(PCS=Process control system)