Compur Statox 4120





Filter cartridge replacement

Remove tube
Remove filter cartridge and bearing
Clean air access opening
Replace filter cartridge and bearing
Replace tube



Generator cell replacement

Disconnect generator cell, remove tube to pump
Remove generator cell

Replace generator cell

Replace tube
Connect generator cell





6.2 Sensor head calibration


6.2.1 Using span gas

What you need:

-Span gas

-Regulator
-Flow

meter

-Gas calibration adapter

-Gas tube (clean, dry, chemical resistant)

-Diagnostic

box


Caution:

The test gas adapter is not part of the ex-certified sensor head.

The gas flow should be adjusted to 500 ml/min (30 l/h).

Statox sensor heads working in the flow mode can be calibrated with the mounted flow adapter.

Keep the working mode gas flow.



Calibration adapter

Flow adapter






23