Compur statox 4120, Technical data, General – Compur STATOX 4120 User Manual

Page 30: Gas measuring range, Sensor head

Advertising
background image

Compur Statox 4120





10. Technical data

General:

Gas Measuring

range

Cl

2

0-1.5 ppm, 0-3 ppm, 0-10 ppm

CO 0-150

ppm

COCl

2

0-0.1 ppm, 0-0.3 ppm, 0-0.5 ppm, 0-1.5 ppm, 0-15 ppm, 0-100 ppm

HCN

0-15 ppm, 0-30 ppm, 0-50 ppm, 0-100 ppm

HCl 0-100

ppm

H

2

S

0-30 ppm, 0-50 ppm, 0-100 ppm

ClO

2

0-0.5

ppm

NO

2

0-15

ppm

SO

2

0-5

ppm



Accuracy at TLV:

+/- 10 %

Alarms:

2 adjustable alarm thresholds, system alarm

RFI: CE

conform

Manufacturer:

Compur Monitors, Munich




Sensor head:

Type:

5330 xxx (xxx: measuring range / gas)

Weight: 1.9

kg

Dimensions (mm, HxBxT):

180x300x180

Operating temperature:

-20 to +40

o

C

Storage temperature:

-30 to +50

o

C

Humidity:

20-95 % r.H. (not condensing)

Pressure: 800-1200

hPa

Protection class:

IP53

Approvals (div.countries):

Cl

2

, COCl

2

, H

2

S

Explosion protection:

EEx ib IIC T6

Operating environment :

II 2 G

Operating voltage U

i

:

max. 22 VDC

Operating current I

i

:

max. 50 mADC

Internal capacity C

i

: 55

nF

Internal inductivity L

i

: 0

mH
















30