Compur statox 4120 – Compur STATOX 4120 User Manual

Page 20

Advertising
background image

Compur Statox 4120





5.1 Using the diagnostic box


Switch the diagnostic box on. Then connect it to the opto- coupler receptacle on sensor head bottom.
Keep the LED`s and the phototransistors of the opto- coupler clean for proper data transmission.


5.1.1 Operation and error codes of the diagnostic box

Note:
If a combination of push button operations is requested ( f. i. ST + ENT), the buttons must be
operated in a short time distance.
Example: ST + ENT initiates a self test. The display says „WAIT-TEST RUN“ until the self test is
finished.


Diagnostic box queries:

Button Display
M

Measured value in ppm

M + ENT

WAIT; after a while reading of actual measured value in ppm

RA

Sensor head measuring range

RA + ENT

Gas

ST

Status passed (System OK) or error message (see 5.1.2)

ST + ENT

(initiates self test), TEST Run

FKT Calibration

factor

ZERO + ENT *

ZERO RUN (start zero adjustment)

* only available with systems with zero adjustment



Diagnostic box error messages:

ERR INPU

Faulty or too slow input. Repeat.

ERR KONZ

Calibration gas concentration out of range.

ERR FACT

Calibration factor is out of range.

DS.W.POS

Measuring range or gas not correctly programmed.

WARN TIM

Time error during data transmission.

WARN DTE

Faulty data transmission.

WARN CHE

No self test due to too low temperature.

TEMP.OU

Temperature out of the admissible range.

TEMP.IN

Temperature is back in the admissible range.

CH.BATT
(audible alarm)

Charge calibration box battery.

NO ZERO

Zero adjustment not possible. Wrong EPROM in sensor
head and control module.

PUMP RUN

Sensor is flushed with air because the gas concentration is
too high.

CAL END

Calibration routine has been cancelled with CE button.

NO CALIB

Calibration not possible within the time limit.