Brookfield R/S Coaxial Cylinder Rheometer User Manual

Page 36

Advertising
background image

Brookfi eld Engineering Labs., Inc.

Page 36

Manual No. M08-219-B1211

This message will be displayed until the fi rst measuring point is displayed:

100.5

000.90 1/s

Eta: 1.728 Pas
Tau: 1.572 Pa
05s 20.7°C

#01

Torque

Duration

Temperature

Actual pre-set value

Measurement result

Current step

If the torque is less than 30‰, the results may be out of instrument range. The user should consider
changing the user defi ned values to get a torque higher than 100‰.

If the display fi eld for the torque indicates: “Mlow!”, these values are below the range of resolution
of the rheometer.

If the temperature is not displayed, the measuring sensor is not connected and the printout of the
temperature will be printed as “---”.

The display is updated with every new measurement point. The current measurement can be

cancelled at any time with the

ST

key.

After a measurement, or after a stop, the display fi eld for step indicates “END” or “BREAK”.
The LCD alternates at intervals of about four seconds between the last displayed measurement
point and information on the measurement:

Program end
Single measurement
Total time: 200s
Total #MP: 40

Reason for

break or end

Duration until

break or end

Type/name of
executed program

Number of
measurement points

Pressing the

OK

key ends the alternating display and returns to the main menu.

VI.5.4 MAIN

Menu

Remote

The “Remote” function initiates measurements to be made under PC control. In this mode, all
functions of the R/S+ Rheometer are controlled by a PC. For PC-controlled measurements, you
need the Rheo 3000 software. This software operates under Microsoft Windows 7™, Windows
2000™, Windows XP™, and Vista™. More detailed information on Rheo 3000 software can be
obtained from Brookfi eld or an authorized dealer.

After selecting the Remote option, the rheometer displays the following:

Remote
Wait for RS232...

<ST>=stop