Brookfield R/S Coaxial Cylinder Rheometer User Manual

Page 57

Advertising
background image

Brookfi eld Engineering Labs., Inc.

Page 57

Manual No. M08-219-B1211

The data in the following table is for plate/plate systems and is valid for a measuring gap of 1
mm. For other plate distances, please calculate K

⋅γ

with the formula:

K

γ

= (2

π/60) * (R/H)

R = radius, H = gap

Table A-4: Plate/Plate Measuring Systems

(consisting of measuring cone and measuring device Cone/plate)

Measuring System

RP3-25

RP3-50

Shear rate range [s

-1

]

1 - 1,309

2 - 2,618

Shear stress range [Pa]

489 - 16,297

61 - 2,040

Viscosity range [Pa•s]

0.373 - 124,500

0.023 - 7,794

Filling volume [mL]

0.5

2.0

Shear rate factor K

⋅γ

[min/s]

1,309

2,6175

Shear stress factor

τ

[Pa]

16,2297

2,037

Radius of measuring plate R

[mm]

12.5

25.0

The given ranges are standard values (not maximum values)

A.2 Error

Messages

The R/S+ Rheometer is user-friendly regarding errors. When errors occur, they are trapped, and the
user is informed on the LCD. The most frequent error messages are explained in this section.

Range Error

Cause:

User’s error at input. The user has tried to input a value that is less than
the allowed minimum value or is more than the allowed maximum value.

Example of
Faulty Input
of a Preset
Value:

Range error
End value:
Min: 0.90
Max: 1032.80

Allowed

minimum value

Name of value out
of range

Allowed maximum
value

What to Do: Enter the value again. Be sure that the new value is within the allowed

value range.