Brookfield R/S Coaxial Cylinder Rheometer User Manual

Page 59

Advertising
background image

Brookfi eld Engineering Labs., Inc.

Page 59

Manual No. M08-219-B1211

BREAK: Torque Max

Cause:

The maximum torque has been exceeded during a measurement.
Possible reasons include:

the measuring range is inappropriate for the sample
the bob cannot rotate in the measuring cup

a]
b]

Example
Message

BREAK:torque max!
Single MODE
total MP: 10
total time: 60

What to Do:

press the OK key

check whether the measuring system has been fastened properly at
the rheometer
if fastened properly, select smaller shear rates or speeds for this
measurement or use a measuring system with a higher shear stress
factor

a]
b]

c]

Abort: Speed Max
BREAK: Speed Max

Cause:

The maximum speed has been exceeded during a measurement. Possible
reasons include:

the selected shear stress or torque value for this measurement is too
high
a torque or shear stress measurement has been started without a
sample in the measuring cup

a]

b]

Example
Message

BREAK:speed max!
Single mode
total MP: 10
total time: 60

What to Do:

press the OK key

select smaller shear stress or torque for this measurement