Brookfield R/S Coaxial Cylinder Rheometer User Manual

Page 58

Advertising
background image

Brookfi eld Engineering Labs., Inc.

Page 58

Manual No. M08-219-B1211

Printer Error

Cause:

The printer has been selected as the output device but is not ready for
printing. Possible reasons include:
a] printer cable is not connected to the rheometer
b] paper is out, printer is not online, other printer errors

Example of
Print Error
Message

Error #1
Printer not ready!

<OK> cont. <ST> stop

Try again

Abort activity;
do not try again

What to Do:

make sure the printer cable is connected to the rheometer
make sure the paper tray has paper
check that the “ready” LED lights of the printer are on

remove the error that was caused at the printer and press the OK
key on the rheometer

press the ST key to abort the process if you cannot fi nd the

error cause

a]
b]
c]

d]

e]

Zero Calibration Error

Cause:

An impermissibly high value was measured during the zero point calibration
of the rheometer.

Example
Message

Error #3
Zero cal. error!
Please retry cal.
<OK> cont. <ST> stop

What to Do:

press the ST key

ensure the measuring system has been removed from the
rheometer
retry zero point calibration - if the error occurs again, this is an
indication of a fault in the measuring instrument

a]
b]

c]

Advertising