Brookfield R/S Coaxial Cylinder Rheometer User Manual

Page 40

Advertising
background image

Brookfi eld Engineering Labs., Inc.

Page 40

Manual No. M08-219-B1211

Select program
1) Prog xyz
2) Test
3) NEW

Available programs

contain a name

All free (non-defi ned)
programs are marked NEW

All free programs are initially marked as NEW. To avoid overwriting existing programs, select
NEW as a program to be edited and select a measuring system. After the selection of the program

with the OK key, you will be requested to select a measuring system.

Attention: Any defi ned program MUST ALWAYS be executed with the same measuring system,
otherwise improper results will be calculated.

Select Meas. system:
1) CC40

>

2) CC25
3) CC14

>

After selecting the measurement system, enter the number of steps (number of ramp and straight
line functions). The number of steps can range from 1 to a maximum of 10 steps. In this
example, we need three steps, so the number “01” is changed to “03” - (see Section VI.4: Input
of Numerical Values
). The message “Range error” will appear if <1 or >10 is entered.

Nr of steps: 03

Input number of
steps 1 - 10. This
example is 03.

After entering the number of steps, input the type of measurement.

Select Input:
-D[1/s]
-n[rpm]
-Tau[Pa] >

Available Measurement Types

User Defi ned Range

Shear Rate D{s

-1

]

Depends on the measuring system

Speed n[rpm]

0.1 - 1,000 rpm

Shear Stress Tau [Pa]

Depends on the measuring system

Torque M[

]

0 - 999 promille (= 0 to 50 mNm)

Select the type of measurement and press the

OK

key (in our example “D[1/s]”). Now enter the

user defi ned values for each step one after the other.