If you change these settings, you also have to set new transfer
parameters of the receiving device!

After starting the function, RS232 Parameters, you will be prompted to enter the baud rate:

RS232: 4800,8,n,1

Change RS232-Par.?
<OK>YES <ST>NO

Press the

OK

key to initiate the input of parameters.

You may enter the following parameters in succession:

• Baud rate [Baud]: 110, 150, 300, 600, 1200, 2400, 4800, 9600, 19200
• Databits [Bit]:

7 or 8

• Parity:

n (=none) e(=even) o (=odd)

• Stopbits

[Bit]:

1 or 2

Select the required parameter from the list of available parameters and accept the selection by

pressing the

OK

key.

Example: Baud rate

Baud rate:
-4800
-9600
-19200

When all parameters have been selected, the new parameters are displayed:

RS232: 19200,8,n,1

<OK>=storing

Accept the set parameters by pressing the

OK

key.

The new settings are now stored and will remain in the rheometer memory even after switching
off the instrument.

VI.7.6 Confi guration

Language

This function selects the user language of the R/S+ Rheometer. Available languages are:

• German
• English