UL-permission and CSA Certifi cate are required for the mains connecting cable.
The following minimal requirements are valid for the cable:

No. 18 AWG
Type SV or similar
3-phase

The cable must have a nominal rating of at least 10 A.
The main plug of type NEMA 5-15P (15A, 125V) or NEMA 6-15P (15A, 250V) must have
a grounding contact.

Japan
The following cable types and connection values are required in Japan:

All parts of the cables (cable, socket and plug) must have brand name and registration
number according to the Dentori law.
The following minimal requirements apply to the cable:

0.75 mm

2

, 2-phase

Type VCT or VCTF
3-phase

The cable must have a minimum nominal rating of 7 A.
The mains plug must be a 2-pin plug with grounding contact according to the Japanese
Industrial Standard C8303 (15A, 125V).

Other countries

The connectors of the mains connecting cable must be approved and certifi ed by the
responsible authorities in the respective countries. These authorities are:

Australia — EANSW

Great Britain — BSI

Austria — OVE

Italy — IMQ

Belgium — CEBEC

Netherlands — KEMA

Denmark — DEMKO

Norway — NEMKO

Finland — SETI

Sweden — SEMKO

France — UTE

Switzerland — SEV

Germany — VDE

The cable must be three-phase HAR-cable, type HO5VV-F3, with a minimal phase
cross-section of 1.0 mm

2

. The main connecting cable must be permitted for a rating of

at least 10 A and, depending on country, a nominal voltage of 125 V or 250 VAC.


-
-
-


-
-
-


-