VIII. Technical Data

R/S+ Rheometer

Dimensions

480 mm x 300 mm x 290 mm

Weight

8 kg

Nominal operating voltage
Power consumption (average)
Power consumption (maximum)

± 15V, 5V
12W
22W

Environmental Conditions
Temperature
in

operation

out of operation

Relative humidity (not condensable)
in

operation

out of operation

10° to 40°C
10° to 45°C

20% to 80%
10% to 90%

Accuracy

± 3-5% of measured value where M>30
± 1 digit

Torque range

1.5 to 50 mNm

Torque resolution

0.01 mNm

Speed range

0.1 to 1,000 rpm

Angle resolution

15.7μrad

Temperature range

-20° to +180°C

depending on the geometry used

Range of shear rate

0.9 to 4 x 10

3

s

-1

depending on the geometry used

Range of shear stress

0.7 to 3.4 x 10

4

Pa

depending on the geometry used

Viscosity range

The given range is a standard value (not maximum value)

1x10

-3

to 3 x 10

3

Pa•s

depending on the geometry used.

Practical low limit is .050 Pa•s for cone/plate measurement.

AC Adapter

Dimensions

160 mm x 85 mm x 35 mm

Weight

0.5 kg

Power supply
Operating

voltage

Output

voltage

Output

current

Output

power

100 to 240 VAC
5V, ± 15V DC
2A, 0.9 / -0.2A
20W

Frequency

50 to 60 Hz

Environmental Conditions
Temperature
in

operation

out of operation

Relative humidity (not condensable)
in

operation

out of operation

+10° to +40°C
+10° to +45°C

20% to 80%
10% to 90%