• This function adapts the electrostimulation

session to the physiology of each user.

Just before starting the work session,

m

i

-

SCAN

probes the chosen muscle group

and automatically adjusts the stimulator

parameters to the excitability of this area of the

body. It is a truly personalised measurement.

N.B.: The

m

i

-

SCAN

function operates as

soon as a stimulation cable fitted with

the

m

i

-

SENSoR

system (sold separately) is

connected to the stimulator.

• This function is implemented at the start

of the programme by a short sequence

during which measurements are taken (a

horizontal bar scans the figurine at the left

of the screen).

• Throughout the entire measurement test,

you must remain completely still and be

perfectly relaxed. The

m

i

-

SENSoR

system is

very sensitive: the smallest contraction or

the slightest movement could disturb the

measurement test.

• Note that certain people might feel an

unpleasant tingling sensation during the

test.

• once the test is complete, the symbol

will appear, indicating that the

programme can be started.

m

i

-

TENS

• The

m

i

-

TENS

function can be used

to considerably reduce the number

of unwanted muscular contractions,

enhancing both user comfort and

treatment efficacy.

• The

m

i

-

TENS

function can only be

accessed for the Acute tendinitis and

Ankle twist programmes.

• For these programmes, the

m

i

-

TENS

function allows you to control the

stimulation energies while maintaining

them at an effective level and

considerably limiting the onset of

muscular contractions.

• Short measurement tests (lasting 2 to

3 seconds) are carried out regularly

throughout the entire programme.

• There is a test phase after each increase

of the stimulation energies. To ensure the

correct performance of these tests, you

must remain perfectly still during this

time.

• Depending on the results of the

measurement tests recorded by the device,

the level of the stimulation energies may

be slightly decreased automatically.

• It is important always to adopt the most

comfortable possible stimulation position.

Also, you should try to remain still and not

contract the muscles of the stimulated

area.

m

i

-

RANGE

• This function indicates the ideal energy

adjustment range for programmes whose

efficiency requires vigorous muscular

twitches.

• The

m

i

-

RANGE

function is therefore only

available in programmes that use low

stimulation frequencies (less than 10 Hz).

• For programmes allowing the use of the

m

i

-

RANGE

function, the stimulator checks

whether you are in the ideal energy range.

If you are below the range, the stimulator

prompts you to increase the energy by
displaying

signs.

• once the stimulator has detected your

ideal adjustment area, a square bracket

will appear to the right of the bar graph

of the channel to which the

m

i

-

SENSoR

system is connected. The figures in this

bracket indicate the energy range you

must work in order to achieve optimal

results.

• If you set the energy level below this

range, the stimulator will prompt you to
increase it by flashing a series of

signs on the screen.