rECommEndations and dECLaration By tHE manufaCturEr -

- ELECtromaGnEtiC immunity

The Compex Runner is designed for use in the electromagnetic environment

stipulated below. The buyer or user of the Compex Runner must ensure it is

used in this recommended environment.

The Compex Runner is designed for use in the electromagnetic environment

stipulated below. The buyer or user of the Compex Runner must ensure it is

used in this recommended environment.

Immunity test

Test level IEC 60601

Observance level

Electromagnetic

environment - recommendations

Electrostatic discharge (DES)

IEC 61000-4-2

± 6 kV at the contact

± 8 kV in air

± 6 kV at the contact

± 8 kV in air

Floors must be wood, concrete or ceramic tiles. If floors are covered with synthetic

material, the relative humidity must be maintained at a minimum of 30%.

Rapid electrical transients in

bursts of noise

IEC 61000-4-4

± 2 kV for power supply lines

± 1 kV for input/output lines

Not applicable

System battery-powered

The quality of the power supply should be that of a typical commercial or

hospital environment.

Surge currents

IEC 61000-4-5

± 1 kV differential mode

± 2 kV common mode

Not applicable

System battery-powered

The quality of the power supply should be that of a typical commercial or

hospital environment.

Voltage trough, short power

cuts and voltage variations

on power supply input lines

IEC 61000-4-11

< 5% VT (trough > 95% of

VT) for 0.5 cycle

< 40% VT (trough > 60% of

VT) for 5 cycles

< 70% VT (trough > 30% of

VT) for 25 cycles

< 5% VT (trough > 95% of

VT) for 5 seconds

Not applicable

System battery-powered

The quality of the power supply should be that of a typical commercial or hospital

environment. If the Compex Runner user requires continuous operation during mains

power cuts, we recommend that the Compex Runner is powered by a UPS or a battery.

Magnetic field at mains

frequency (50/60 Hz)

IEC 61000-4-8

3 A/m

Magnetic fields at the mains frequency should be at the level of a representative site

located in a typical commercial or hospital environment.

notE: Vt is the aC supply voltage before application of the test level.