9. Problems and solutions

Electrode fault

The Compex bleeps and alternatively

displays the symbol of a pair of electrodes

and an arrow pointing to the channel

where a problem has been detected. In

the example above, the stimulator has

detected an error in channel 1.

• Check that electrodes are connected to

this channel.

• Check whether the electrodes are old,

worn, and/or the contact is poor: try

using new electrodes.

• Try using the stimulation cable on a

different channel. If the cable is still

showing a fault, replace it.

Stimulation is not producing the

usual sensation

• Check that all the settings are correct

and ensure the electrodes are positioned

properly.

• Change the positioning of the electrodes

slightly.

The stimulation effect causes

discomfort

• The electrodes are beginning to lose

adhesion and no longer provide good

contact on the skin.

• The electrodes are worn and need to be

replaced.

• Change the positioning of the electrodes

slightly.

The stimulator is not working

• If an error screen appears while the

equipment is in use, note the error number

(in the example in fig. 15, error 1/0/0) and

contact the consumer service stated and

approved by Compex Médical SA.