Hardy HI 1756-nDF User Manual

Page 41

Advertising
background image


37

Chapter 4

Parameter 23

Stop refill weight

Weight at which a refill cycle will automatically stop
Value must be greater than 0.0 (default) and equal to or less than the stop refill weight
4000.0 (default)

Parameter 24

Capacity

When the gross weight is greater than the programmed capacity value the
ERROROVERCAPACITY status bit is set. This status bit is reset whenever the gross
weight is below programmed capacity value.

Parameter 25

Flow rate time base

The Time Base is the length of time in seconds between two weight readings that are
subtracted to determine the initial flow rate. By increasing the time base, you increase the
time between weight readings. This allows more material to be dispensed during the time
base period. Low flow rates require a longer time base than high flow rates. The
controller measures weight to about 1 part in 10,000. The formula below provides a
minimum.

TIME BASE > (SCALE CAPACITY/10000)/LOW SETPOINT (units per sec.)

SCALE CAPACITY/5000 = determines minimum weight increment

Parameter 26

Reserved.

Parameter 27

variation

Sets the peak to peak variance value below which samples are filtered.

Parameter 28

capture

Sets the peak to peak variance value used to select which filter is used to provide the
output value.

Parameter 29

start wait

Period of time, from the start of the fill cycle, during which the flow rate is not
calculated.

Parameter 30

Update Time

Period of time, after the target weight has been reached before the measured weight value
is taken.

Parameter 31

stable

Sets the maximum peak to peak percentage difference in flow rate, once the “stable wait’
period is reached. Values outside of this range reset the calculation.