Chapter 4

Commands for weigh scale 2 from the PLC, output table offset 8 – 15, are only valid for
HI 1756 2DF modules. For the HI 1756 1DF modules this portion of the dynamic output
table will be ignored.

Status Word
The “status word” is a bit encoded value within the integer number.

Status Word

Value

Definition

STATUSCHANENABLED

0x00800000

Set if channel enabled

STATUSINMOTION

0x00400000

Variations in weight exceed the
motion tolerance parameter

ERRORADCONVERT

0x00200000

Millivolt return from the load cell
system is out of range for the unit

ERRORADFAILURE

0x00100000

Unit’s A/D converter not
responding.

ERROROVERCAPACITY

0x00080000

Out of range value

ERROR_EPROM

0x00000400

EPROM hardware error

STATUS_AP_WT_NOT_STEADY

0x00000100

Indicates if the current sample is
within tolerance.

STATUS_AP_NOT_STABLE

0x00000080

Set if flow rate was not stable.
Remains set until the start of the
next fill cycle

STATUS_AP_NOT_STEADY

0x00000040

Set if weight on the scale was not
within tolerance at the end of the
fill cycle. Remains set until the
start of the next fill cycle

STATUS_AP_WAITING

0x00000020

Set after the last relay closes and
remains set until the weight
sample is taken

STATUS_AP_ACTIVE

0x00000010

Set during the entire fill cycle, if
auto preact enabled

STATUS_REFIL

0x00000008

Refill process active

STATUS_ON

0x00000004

System ON or OFF

RELAYB_ASSERTED

0x00000002

Relay B is asserted

REALYA_ASSERTED

0x00000001

Relay A is asserted

*Dual Mode and auto refill ON, or in the auto refill mode where the target weight greater than the
difference between refill low and refill high or trying to start while the scale is in motion.

Integrated Technician

If there is NO IT JBOX, the instrument can still measure the sense voltage, the overall
load cell millivolts per volt, and the internal reference weight value. If the user has an IT
JBOX, the instrument can also read the weights and voltages of the individual sensors
attached to the IT JBOX.

To initiate an IT test; do a WRITEPARAM command, with parameter number 0x0036,
and a parameter value equal to the number of sensors, which should be 1-4 if you have an
IT JBOX. If you do not have an IT JBOX, the parameter value does not matter.

To read back the results, do a READPARAM command, with parameter numbers
0x0036, 0x1036, 0xB036. The return values are listed below.