Chapter 6

The load cell/point uses the 5 volts DC excitation voltage generated by the HI 1756 nDF.
It generates a millivolt output proportional to the weight on the scale (0-10mV DC for
2mV/V load cells/points or 0-15mV DC for 3mV/V load cells/points).

The HI 1756 nDF instrument provides among other functions:

Power to the load cell(s)/point(s)

Reads the millivolt signal output from the load cell(s)/point(s)

Digitizes, interprets, communicates the results in terms of weight and rate

INTEGRATED TECHNICIAN (IT®)

INTEGRATED TECHNICIAN (IT®) is a diagnostics utility that enables the operator to
rapidly troubleshoot the individual load cells in a weighing system. IT commands are
communicated via the C2 wires connecting the summing card and controller. You must
use C2 certified cable, with or without C2 load sensors. Insure the Violet and gray wires
are properly installed. An HI 215IT Summing junction box is required to read data for
individual load points. It provides separate inputs for up to four load cell.

If you have more than one load cell without the IT junction box, there is no easy way to
isolate the load cell signals.

If you have more than one load cell with the IT junction box, the system can provide both
the average numerical values and values specific to each load cell. The number for a load
sensor is based on the connections in the IT junction box.

Check the installation sequence in the box to determine which load sensor is number 1, 2
and so on. Always install the first load cell in position 1.

If you do not have the optional IT junction box, some of the options described below will
not appear on either interface.

JUNCTION

BOX

INTERMEDIATE SUPPORT FRAME

FULLY CONSTRAINED

LATERALLY WITH STAY RODS

UNIVERSAL JOINT

OR

HOSE

USE SYMMETRIC BEAM LOADED

THROUGH SHEAR CENTER TO

AVOID TWIST WITH LOAD

NON METALLIC EXPANSION

ASSEMBLY OR HOSE

BIN ACTIVATOR