F - verify individual load cell millivolt readings – Hardy HI 1756-nDF User Manual

Page 57

Advertising
background image


53

Chapter 6

F - Verify Individual Load Cell Millivolt Readings

Testing an individual load cell signal output requires an IT Summing Junction box or
millivolt meter.

Use the load cell certificate to verify the millivolt per volt (mV/V) rating:

Example: 3mV/V load cells produce approximately 15mV at full load. That is 5 volts
excitation x 3 mV/V. At a scale capacity of 1,000 lbs. with 100 lbs. of deadload when
empty, the load point mV reading should measure 1.5mV.

F

Select

IT

TEST

MV/V

IT TEST Checks

individual load

sensor output

mV/V readings

Select

load

sensor

Record load

sensor mV/V output

level for

comparison

Repeat for all load

sensors

MV/V

readings

acceptable?

Zero mV/V

reading

1) No dead load

Apply load and re-test.

2) Wiring error

Verify color code using the

load cell certificate

3) Open bridge circuit

Disconnect power and verify

load point bridge resistance

reading with an Ohmmeter

1) Stressed load cell, remove

all load and re-test.

2) Excessive loading

For additional

testing go to:

H

High mV/V

reading

K

Defective load Cell

Replace and repeat

TEST F

TEST COMPLETE

Yes

No

High mV/V

reading