2 method setup, 1 method setup - acquisition – Clarity INT9 User Manual

Page 21

Advertising
background image

INT9 A/D converter

4 Using the INT9 card

4.2 Method Setup

Parameters of the

Method Setup - Acquisition

and

Method Setup -

Measurement

dialogs depend on the type of analysis (measurement

conditions), so they can be specific for various types of analysis. This is
the reason for them to be a part of the method.

Caution:

The parameters can't be changed during analysis.

4.2.1 Method Setup - Acquisition

Fig 13: Method Setup - Acquisition (INT9)

Select Detector

Selects the detector whose parameters will be set. If there is only one
detector (channel) configured on the Clarity Instrument, this control will
not be accessible.

Range

Selects the input voltage range (in mV). The selected range corresponds
to the maximum output voltage of the detector that the A/D converter can
still process.
Available ranges for INT9 card are 156, 1 250 and 12 123 mV.

Sample Rate

Depending on Supply Frequency filter selected in the

System

Configuration

DataApex INT9 Setup

dialog, see the chapter "DataApex

INT9 Setup" on pg 13 . The available sampling rates are 6.25, 12.5 , 25 ,
50, 100 , 200 and 400 samples per second (for 50Hz supply frequency

- 17 -