3 Installation

Clarity Hardware

a set of one signal wire, one starting wire and one digital output wire. The
2-channel INT9 card would thus have the cable with following parts:

Signal cables

Labeled DET1 and DET2 , the cables are supplied as standard without
connectors [bare stripped and tinned endings – red (+), white (-) and
shielding (analogue ground)].

Starting (marker) cables

Labeled IN1 and IN2 , ended with CINCH connector (female). The
following cables can be connected to this CINCH connector:

l

A cable terminated with free leads [red (+), shielding (digital ground)] for
direct connection to the chromatograph.

l

A cable terminated with a button for cases where a starting contact is not
available and it is necessary to perform a manual start.

Both types of cable are supplied for each starting cable.

Digital output cables

Relay contacts labeled OUT1R and OUT2R, terminated with free leads.
These are used for synchronizing autosamplers without an AS Control
module in the ACTIVE Sequence or for controlling other devices.
At the converter end the cable is always terminated with a DB- 37F
connector (37-pin female connector).