1 int9 acquisition card, 1 a/d converter types in general – Clarity INT9 User Manual

Page 5

Advertising
background image

INT9 A/D converter

1 INT9 acquisition card

1 INT9 acquisition card

This manual describes the use of the INT9 A/D converter with the Clarity
software ver. 2.7 and later.
INT9 is an internal 24-bit A/D converter card that measures the voltages of
chromatographic and electrochemical detectors.

Fig 1: INT9

Brief technical overview

l

Up to 4 dataacquisition channels on one card (one and two channel
versions are also available).

l

The channels are completely independent sigma delta integrating A/D
converters, used for processing positive and negative voltages.

l

4 digital inputs.

l

8 digital outputs. The first four are also designed as relay contacts.

l

PCI Slot.

l

Excellent reliability and temperature stability.

The INT9 card of integration converters is a modernized successor of the
PCI card INT7.

1.1 A/D converter types in general

Prior to collecting and processing the chromatographic signal by a
computer, the analog signal of the detector must first be converted to
digital form using an analog/digital (A/D) converter.
In principle, there are three types of A/D converters – sampling,
integrating and δΣ (Delta-Sigma) - with continuous integration). The INT9
card uses Delta-Sigma integration.

- 1 -