INT9 A/D converter

4 Using the INT9 card

4.2.2 Method Setup - Measurement

Fig 14: Method Setup - Measurement

External Start/Stop

Enables control from an external signal.

Note:

The input used for an external start from a chromatograph can be set in
the

System Configuration

dialog, see Fig 4 on pg 8.

Start Only

The external signal only starts an analysis. Triggering the input will not
influence the acquisition in progress.

Start - Restart

The external signal terminates the analysis in progress and at the
same time starts a new one. The first signal starts the analysis; the
second signal terminates it and at the same time starts a new one - the
so-called continuous measurement.

Start - Stop

The external signal starts and stops an analysis. The first signal starts
the analysis; the second signal stops it, etc.

Down

The station reacts to changes in voltage in relevant controlling input
from high (> 3 V) to low (< 0.7 V), or to contacts closed in the relay.

Up

The station reacts to changes in voltage in relevant controlling input
from low (< 0.7 V) to high (> 3 V), or to contacts opened in the relay.