background image

6 Tables and Specifications

Clarity Hardware

6.4 Technical data

Tab 5: Technical data of the INT9 board:

Parameter

Description

Converter type:

24-bit δΣ (delta – sigma)

Analog inputs:

differential

Number of
channels:

1, 2 or 4 fully independent and isolated

Input range:

unipolar and bipolar, each channel independently 156, 1 250
or 12 123 mV

Input impedance:

> 1 MΩ

Sample rate:

6.3, 12.5, 25, 50, 100, 200, 400 samples per second for 50Hz
mains frequency filter
7.5, 15, 30, 60, 120, 240, 480 samples per second for 60Hz
mains frequency filter

Non-linearity:

< 0.003 %

Temperature
dependence:

< 10 ppm/°C

Digital inputs:

4 TTL (5 V)

Digital outputs:

8 TTL (5 V, maximum current load per output 20 mA,
maximum total current load 100 mA), the first four control
relays (switching contacts rating: max. switching current 0.5
ADC, max. contact resistance 150 mΩ, max. voltage 100 V)

Tab 6: Typical noise free resolution for various acquisition speeds and input ranges:

12 123 mV

1 250 mV

156 mV

6.3 Hz

21 bit

21 bit

20 bit

25 Hz

21 bit

21 bit

20 bit

100 Hz

19 bit

19 bit

19 bit

200 Hz

16 bit

16 bit

16 bit

400 Hz

16 bit

16 bit

16 bit