ADLINK PXI-2022 User Manual

Page 29

Advertising
background image

Signal Connections

21

3.2.2

Input Connect Configurations - Differential Input

Mode

Figure 3-1: Ground-referenced Source and Differential Input

The differential input mode provides two inputs that respond to
signal voltage difference between them. If the signal source is
ground-referenced, the differential mode can be used for the com-
mon-mode noise rejection. Figure 3-1 shows the connection of
ground-referenced signal sources under differential input mode.

Figure 3-2 shows how to connect a floating signal source to the
PXI-2020/2022 card in differential input mode. For floating signal
sources, you need to add a resistor at each channel to provide a
bias return path. The resistor value should be about 100 times the
equivalent source impedance. If the source impedance is less
than 100ohms, you can simply connect the negative side of the
signal to AIGND as well as the negative input of the Instru-menta-
tion Amplifier without any resistors. In differential input mode, less
noise couples into the signal connections than in single-ended
mode.

AIxH

AIxL

x = 0, ..., 31

Ground
Referenced
Signal

Source

To A/D

Converter

AIGND

+
-

Instrumentation
Amplifier

V

cm

Common-

mode noise &
Ground

potential

+

-

Input Multiplexer

Advertising
This manual is related to the following products: