Function Block and Operation Theory

4.3

ADC Sampling Rate and TIMEBASE Control

The PXI-2022 supports six timebase sources for analog input con-
version:

1. On board Internal oscillator

2. External clock through front panel (AFI[0..7])

3. External clock through front panel SMB CLK IN

4. PXI Star Trigger

5. PXI Trigger Bus Line 0

6. PXI 10M

The following diagram shows the timebase architecture of the PXI-
2022.

Figure 4-4: PXI-2022 Timebase Source and Architecture.

4.3.1

Internal Oscillator

The PXI-2020/2022 equips a high stability, low jitter oscillator for
the ADCs. The oscillators are 80 MHz for the PXI-2020/2022.

4.3.2

External Clock through Front Panel

When you need a specific timebase in some applications that the
onboard oscillator is not achievable, a clock from an external
device can replace onboard oscillator. In addition, external time-
base also provides a method to synchronize the DAQ module to
other measurement modules by distribut-ing/receiving a common
clock to/from multiple modules. The PXI-2020/2022 can receive

T

im

eb

ase

C

lo

ck

M

ux

P

X

I Inter

face

PXI Trigger Bus Line 0

PXI_STAR

Ext. CLK IN

SMB

Connector