Labconco Glassware Washer Service/Technical Manual

5

Inlet Solenoid
Valves

(Blue) Top

(rear)

Left

Electrically controlled to
allow water to fill the tank
of the machine.

DI/Pure Water Pump (Pink)

Bottom(front)
Left

Integrated pump to allow the
use of a non-pressurized Pure
water source.

Wash Pump/Motor

(Yellow)

Top (rear)
Right

The pump has been moved away
from the sump in order to

minimize the occurrences of
motors being damaged by
leaks.

Drain Pump &
Solenoid

(Red) Bottom(front)

Right &
Center

When the machine sets to
drain, the Solenoid will
“pop” open, three seconds
later the drain pump will

kick on and the water will
leave the tank.

Drying Blower

(Light
Green)

Left Center

The blowers are different
between the FlaskScrubber®

and SteamScrubber®. These

blowers circulate air through

the tank to dry labware after
it has been washed.

Overflow Sensor

(Dark

Green)

Top (rear)

Center

This sensor is a set of

contacts that when wet have

continuity, signaling the

washer that there is water in

the base weldment pan.

FlaskScrubber®

SteamScrubber®

These illustrations are to be compared with the photos on the previous page
and are in the same orientation with the colored circles representing the
same parts. The Purple Circle in the middle represents the Pump Outlet
Elbow.