Labconco Glassware Washer Service/Technical Manual

24

Hose from washer must not be
attached to another hose.

Check for kinks in hose.


Drain hose must not be elevated more
than 30 inches.

Pump lost its prime

Remove ALL water from inside drain
hose. Pump will prime when water

drains from washer.

Building plumbing

defective

Check for blocked air gap.

Check for clogged drain.

Filter screen
clogged

Clean screen in bottom of washer.

Defective fill level
Switch

Contact Labconco Product Service.

Defective controller Check output from PCB to drain valve

and pump.

Replace PCB. Contact Labconco
Product Service.

Water High Alarm

Holes in filter
screen too small

Contact Labconco Product Service.

Defective hose to

fill switch

Check for splits.

Check for loose clamps.

Defective fill level
switch

Recalibrate switch. Contact Labconco
Product Service.

Replace switch.


Temp Sensor Alarm

Holes in filter

screen too small

Poor Connection

between Sensor and

PC Board





Contact Labconco Product Service.

Check Connectors between Temp Sensor

and Sensor Wiring Harness.

Inspect integrity of Sensor Wiring

Harness.

Check Connectors between Sensor

Wiring Harness and Relay PC Board.


Check Connectors between Relay PC
Board and Ribbon Cable.


Inspect integrity of the Ribbon
Cable.

Advertising