Labconco Glassware Washer Service/Technical Manual

25



Faulty Temperature
Sensor


Check Connectors between Ribbon
Cable and PC Board.


Temp Sensor Bad. Replace.

Water Low Alarm

Improper water

supply

Make sure water is turned on.

Tap water pressure must be at least
20 psi (138 kPa) and flow must be at
least 1.25 gallons (4.7
liters)/minute.


DI water supply must have flow of at
least 0.9 gallons (3.4

liter)/minute.

At least 3.4 gallons (13 liters)
must be available for each DI rinse.

Defective fill
valves

Clean screens inside the inlet
fittings of the water valves.

Check voltage to valves.

Replace valve if it does not
operate.

Restricted flow

Check for kinks in water hoses in

base of washer to water inlet

fitting on side of tank.

Defective DI water

pump

Check voltage at pump.

Listen for pump operation. Replace

if necessary.

Defective hose to

fill switch

Check for splits.

Check for loose clamps.

Defective fill level

switch

Recalibrate switch. Contact Labconco

Product Service.

Replace switch.

Defective controller Check output from PCB to both valves

and DI pump.

Replace PCB. Contact Labconco

Product Service.

Holes in filter

screen too small

Contact Labconco Product Service.

Defective grommet

Replace grommet.