62

INSP/TESTS No. CONDITIONS

AND

REQUIREMENTS

INST/LCC# RESULTS PASS

115 V 230 V

DETERGENT CUP

Audible

HEATER 9.3 – 11.5 Amp 7.8 – 9.6 Amp

Ammeter __________

Return to TAP – Allow water to fill washer until it turns off. Press and
hold RUN – Water must turn off before it overflows.

Visual

DRAIN / COOL – Allow all water to drain

Visual

DRY

BLOWER

Audible

LIQ DETERG PUMP (No test)

LIGHT (Window models only)

Visual

FAN

Aubile

Run Test

1

Fill Detergent Cup with Detergent. Close Door

2

Scroll

through

Programs

Confirm Display Information

Visual

3

115V Models – Select Glass Plus 230V Models- Select Glass Plus
Wash 2 Temp 82°C Wash 2 Temp 93°C
Wash 2 Time 10 min Wash 2 Time 10 min
Steam Yes Steam Yes
No. Rinses 3 No. Rinses 3
No. Pure Rinses 1 No. Pure Rinses 1
Final Rinse Temp 60°C Final Rinse Temp 60°C
Dry Temp 60°C Dry Temp 60°C
Dry Time 20 minutes Dry Time 20 minutes
Start Data Acquisition Computer – Identify file with washer serial
number and date

Press RUN

Confirm operation of washer, temperatures and times
Acceptance Criteria:

Visual-

RS232

Wash 2 and Final Rinse Temp +/- 3°C

Dry Temp +/- 8°C once reached

Wash 1, Steam, Rinses 1 & 2, Dry Time +/- 1 minute

Wash 2, Final Rinse Time after reaching SP +/- 1 minute

4

All Window Models

Observe Wash Action

FlaskScrubbers – Water Hits Top of Tank from spindles

Visual

SteamScrubbers – Lower Wash – Arm Rotates at least 45 RPM

Stop Watch

5

Run rinse only to wet surfaces.

Visual

After wash pump starts, place a wet towel across the overflow
contacts. Wash pump should stop and the washer should drain.
Display should show ALARM OVERFLOW.