Glassware Washer/Test Report

MODEL NUMBER

VOLTS/HZ

M.O. NUMBER

SERIAL NUMBER

INSP/TECH/

DATE

QTY/OF

INSP/TESTS No. CONDITIONS

AND

REQUIREMENTS

INST/LCC# RESULTS PASS

Set-Up

1

Connect tap and pure water.

2

Connect

drain.

3

Connect electrical power. (230V, 50 Hz)

4

Connect current probe to power line.

5

6
7
8

Close detergent cup door
Connect computer via RS232
Connect test dispense bottles to the detergent & acid supply lines.
Set up washer control
A. Open door
B. Press and hold ▼ button- Simultaneously close and latch door.
C. Press ▼ button to select function. Press ◄ or ► to actuate
function.
1) Set Temp Units - °C
2) Set DI Pump to Enable- YES
3) Set Light - ON
4) Set Factory Reset – YES

D. Press RUN

Hi-Pot

1

1000V for 1 minute.

Hi Pot

Tester ____________

Temperature 1

Position thermocouple near temperature probe.

Test

2

Press and hold ◄ button - simultaneously close and latch door.

3

Display must agree with temperature meter

±2°C.

Temperature

Meter

Self Test

1

Open

door

2

Press & hold ▲ button – Simultaneously close and latch door.

3 Press

▼ button to select function. Press ◄ or ► to actuate

function.

TAP

WATER

Visual

PURE

WATER

Visual

Continue in this mode until Float Switch shuts off water flow.

Visual

Open door to check water level. Bottom of heater must be