Labconco SteamScrubbe & FlaskScrubbe Glassware Washers 44204 Series User Manual

Page 29

Advertising
background image

Labconco Glassware Washer Service/Technical Manual

23

Conductivity
reading “0”

Sensor not connected Reconnect sensor wires.

Sensor defective

Replace sensor.

Sensor dirty

Clean sensor.

Conductivity
Reading

fluctuates

Temp Reading
Stuck or
Fluctuating at

16ºC

Washing action cause
Bubbles



Poor Connection

between Sensor and
PC Board

None.



Check Connectors between Temp Sensor
and Sensor Wiring Harness.

Inspect integrity of Sensor Wiring
Harness.

Check Connectors between Sensor
Wiring Harness and Relay PC Board.


Check Connectors between Relay PC
Board and Ribbon Cable.

Inspect integrity of the Ribbon

Cable.

Check Connectors between Ribbon
Cable and PC Board.

Drain Fail Alarm

Drain Valve

inoperable

Listen for click when washer should

drain.

Check for mechanical restriction of

actuator arm on top of valve.

Check for electrical power to valve.

Check for damage to valve coil.

Drain Pump

inoperable

Check if pump fan turns when washer

should drain.

Check for mechanical restriction of

fan.

Check for electrical power to pump.

Drain hose not

routed
properly

Hose from drain pump must slope

upward and have no low spots or

horizontal runs under washer tank.

Make sure hose is attached to the

back of the washer and has a high
loop.

Hose must be attached to building
plumbing pipe ¾ inch ID or larger.

Advertising