KB (E3.1 + E5.1) 11/2014

page 33/100


Program entry as set-point ramp (example):

W/°C

t/min.

S01

S02

S03

S04

0

20

40

60

80

100

30

120

180

380

S05


Program table corresponding to the diagram (with default setting “Ramp”):

Section


SEC

Temperature

set-point

[ °C]

TEMP

Section length

[hh.mm]

TIME

Fan speed

[%]

FAN

Operation lines

*


O.LINE

S01

40

00:30

50

000

S02

60

01:30

100

000

S03

90

01:00

100

000

S04

90

03:20

100

000

S05

20

00:01

100

000

* Only with option zero-voltage relay outputs via operation lines, see chap. 16.4.
The values of such a program table can now be entered to the RD3 program controller (chap. 8.2).

Program entry as set-point step (example):

W/°C

t/min.

S01 S02 S03 S04 S05 S06 S07 S08

0

20

40

60

80

100

30

120

180

380