Fault description

Possible cause

Required measures

Refrigerating performance

Low or no refrigerating

performance.

Ambient temperature above

25 °C / 77°F (chap. 3.4).

Select cooler place of installation.

Compressor not turned on.

Contact BINDER service.

Electro-valves defective.
No or not enough refrigerant.
Too much external heat load.

Reduce heat load.

Controller
Program duration longer than

programmed.

Inappropriate tolerances have

been programmed.

For rapid transition phases, do NOT

program tolerance limits in order to

allow maximum heating speed.

Program stops one section too

early.

Program line is incomplete.

When programming, define the end

value of the desired cycle by adding

an additional section with a section

time of at least one minute (with

setting “ramp”).

Programs have been deleted.

Change from 2 programs to 1

program or vice-versa

When changing, ensure that the

programs are no longer needed.

The controller returns to Normal

Display from any level.

No button was pressed for more

than 120 sec.

Repeat entries, enter the values

rapidly.

Message RANGE ERROR CH1

in Normal Display in Display 2

Sensor rupture between sensor

and controller, or Pt 100 sensor

defective.

Contact BINDER service.

Ramp temperature transitions

are only realized as steps.

Set-point programming type set

to “Step” in the User level (chap.

10).

Set the set-point programming type

to setting “Ramp” in the User level

(chap. 10).

Only qualified service personnel authorized by BINDER must perform repair.

Repaired units must comply with the BINDER quality standards.