Danger – BINDER KB 23 User Manual

Page 74

Advertising
background image

KB (E3.1 + E5.1) 11/2014

page 74/100


Maximum loading capacity of the switching contacts: 24V AC/DC – 2.5 A

DANGER

Electrical hazard.
Danger of death.
Damage to switching contacts and connection socket.

Do NOT exceed the maximum switching load of 24V AC/DC – 2.5A.

Do NOT connect any devices with a higher loading capacity.

16.7 Additional flexible Pt 100 temperature sensor (option for KB 53, 115)

An additional flexible Pt100 temperature sensor allows measuring the temperature of the charging

material by means of an independent measuring system utilizing Pt 100 entry. The Pt 100 sensor’s top

protective tube can be immersed into liquid substances.
Technical data of the Pt100 sensor:

Three-wire technique

Class B (DIN EN 60751)

Temperature range up to 320 °C / 608 °F

Stainless steel protective tube with a length of 45

mm / 1.78 in, material no. 1.4501





Lemo socket

Lemo plug

Figure 22: Option Pt 100 temperature sensor

16.8 Additional flexible Pt 100 temperature sensor (option for KB 240, 400, 720)

An additional flexible Pt100 temperature sensor allows measuring the temperature of the charging

material by means of an independent measuring system utilizing Pt 100 entry. The Pt 100 sensor’s top

protective tube can be immersed into liquid substances.

Pt 100 Sensor
PIN 1: Pt 100
PIN 2: Pt 100
PIN 3: Pt 100





DIN socket (9)

DIN plug


A suitable DIN plug is enclosed.

Figure 23: Pin configuration of the DIN socket (9) in the right lateral control panel

Technical data of the Pt100 sensor:

Three-wire technique

Class B (DIN EN 60751)

Temperature range up to 320 °C / 608 °F

Stainless steel protective tube with a length of 45 mm / 1.78 in, material no. 1.4501

1 2 3 4 5 6

1 2 3 4 5 6

1 2 3

1 2 3

2 3

1

Advertising